Techmize TH600 Comprehensive Cable Testing System

Product number: TH630-640
The Techmize TH2838 LCR meter offers 20 Hz–2 MHz with auto-balance bridge technology and 0.05% accuracy.

Price on request

Model
Request a quote
Paypal-icon
Mastercard-icon
visa-icon
klarna-icon
Sepa-lastschrift-icon
invoice-icon
  • 2-year warranty
Further information

Description

The TH6XX series are modular wire/cable and cable harness testers with a 7″ TFT-LCD display, integrated automation interface, and versatile test procedures. Different models (TH610, TH615, TH630) offer varying numbers of channels and voltage ranges for scalable test setups in production and laboratory environments.

Features

  • 7″ True-Color TFT LCD (800×480) with Chinese/English GUI and touch/keyboard operation 
  • Modular design with support for 2-wire/4-wire test conversion, up to 5120 scan channels 
  • Continuity, short circuit and instant short/open circuit tests with automatic point search 
  • AC/DC mid-voltage tests with arc detection and high/low voltage separation (insulation >100 GΩ) 
  • Component testing (resistance, capacitance, diode, polarity) using four-wire technology and parallel voltage/current sensing 
  • Sequence and automation functions: Pass/Fail outputs, HANDLER interface, RS232/RS485, USB Host/Device, LAN, GPIB options 
  • Thermistor measurement (NTC) and contact testing before the start of testing 
  • Instrument self-test, online diagnosis, firmware upgrade via USB stick 
  • Robust design for use in automation systems, production test benches and laboratory environments 

Applications

    • Automotive electronics: battery, ECU and sensor wiring, cable harnesses, audio/video connections 
    • Communication/IT: Telephone, network, USB/HDMI/VGA/SATA cable tests 
    • Electronics industry: flat ribbon cables, connectors, power cables, multi-core sockets 
    • Component testing: Passive components (capacitor, inductor, resistor, diode, polarity) 
    • Safety testing: AC/DC insulation and resistance tests for cables and materials 
    • Automated production test benches: sequence tests, pass/fail evaluation, data acquisition, PC control (SCPI/Modbus) 
    • R&D & Laboratory: Prototype testing, long-term diagnostics, automated network quality testing

Technical Data

Model comparisonTH610-2560
TH610-2560
TH610-5120
TH610-5120
TH615-640
TH615-640
TH615-1280
TH615-1280
TH630-320
TH630-320
TH630-640
TH630-640
Scan channels 4-wire 1280 2560 640 1280 320 640
Scan channels 2-wire 2560 5120
Max. Module 20 40 10 20 10 20
DC source 0.1–5 V; 2% at idle 0.1–5 V; 2% at idle 0.1–5 V; 2% at idle 0.1–5 V; 2% at idle 0.1–5 V; 2% at idle 0.1–5 V; 2% at idle
DC power source 1–15 mA; 2% 1-20mA, 100mA or 1A; 2% 1-20mA, 100mA or 1A; 2% 1-20mA, 100mA or 1A; 2% 1-20mA, 100mA or 1A; 2% 1-20mA, 100mA or 1A; 2%
Program. HV source DC 50–1000 V DC;
±(1% of the target value + 0.5% of the final value)
50–1000 V DC;
±(1% of the target value + 0.5% of the final value)
50–1500 V DC;
±(1% of the target value + 0.5% of the final value)
50–1500 V DC;
±(1% of the target value + 0.5% of the final value)
50–3000 V DC;
±(1% of the target value + 0.5% of the final value)
50–3000 V DC;
±(1% of the target value + 0.5% of the final value)
Program. HV source AC 50–750 Vrms;
±(1% of the target value + 0.5% of the final value)
50–750 Vrms;
±(1% of the target value + 0.5% of the final value)
50–1000 Vrms;
±(1% of the target value + 0.5% of the final value)
50–1000 Vrms;
±(1% of the target value + 0.5% of the final value)
50–2000 Vrms;
±(1% of the target value + 0.5% of the final value)
50–2000 Vrms;
±(1% of the target value + 0.5% of the final value)
AC signal source 50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
50 Hz–100 kHz;
1 Vrms; 0.02% frequency accuracy
Capacity measurement 10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
10 pF–1000 µF;
±(1 % of the measured value + 0.005 % of the final value)
Resistance test 10 mΩ–1 MΩ 10 mΩ–1 MΩ 1 µΩ–1 MΩ 1 µΩ–1 MΩ 1 µΩ–1 MΩ 1 µΩ–1 MΩ
Short circuit/open circuit test 1 kΩ–50 kΩ; 10% 1 kΩ–50 kΩ; 10% 1 kΩ–50 kΩ; 10% 1 kΩ–50 kΩ; 10% 1 kΩ–50 kΩ; 10% 1 kΩ–50 kΩ; 10%
Diode test 0–10 V;
±(5% of the measured value + 1% of the final value)
0–10 V;
±(5% of the measured value + 1% of the final value)
0–10 V;
±(5% of the measured value + 1% of the final value)
0–10 V;
±(5% of the measured value + 1% of the final value)
0–10 V;
±(5% of the measured value + 1% of the final value)
0–10 V;
±(5% of the measured value + 1% of the final value)
Insulation resistance 1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
1 MΩ–1000 MΩ;
±(3% + 0.1%) or ±(7% + 2%)
DC leakage current 1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
1 µA–5 mA;
±(5% of the measured value + 0.04% of the final value)
AC leakage current 0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
0.01–5 mA;
±(10% of the measured value + 0.1% of the final value)
Contact test Only in 4-wire mode Only in 4-wire mode Only in 4-wire mode Only in 4-wire mode Only in 4-wire mode Only in 4-wire mode
HANDLER interface 8 outputs and 8 inputs 8 outputs and 8 inputs 8 outputs and 8 inputs 8 outputs and 8 inputs 8 outputs and 8 inputs 8 outputs and 8 inputs
Dimensions & Weight 483 × 396 × 535 mm;
approx. 35 kg
483 × 660 × 535 mm;
approx. 60 kg
483 × 396 × 535 mm;
approx. 35 kg
483 × 660 × 535 mm;
approx. 60 kg
483 × 396 × 535 mm;
approx. 35 kg
483 × 660 × 535 mm;
approx. 60 kg