Datalogging with four-channel SDS1000X-E oscilloscopes

Data collection

Many test and measurement applications require data collection over a long period of time. Instead of milliseconds of measurement time, you may need tens of minutes or even hours of data. A data acquisition system is typically used for this type of application. However, this may require the development of software and debugging code. SIGLENT has added a data logger mode to the four-channel oscilloscopes of the SDS1000X-E series, which makes the acquisition of long-term data easier than ever. The SIGLENT data logger continuously records the original sampling points and measured values. The Siglent SDS1000X-E includes the devices SDS1104X-E, SDS1202X-E and the SDS1204X-E oscilloscope.

XEDatalogger_1

Setting the system time

The SIGLENT SDS1000X-E oscilloscopes do not have a system clock. If you want the data logger time to be the correct time, you need to set the system time after power on. You can set the system time manually or configure the network time server address to get the time from the network time server via NTP as shown below:

XEDatalogger_Date_Time-2

The sample logger

 The Sample Logger continuously records and saves measurement points. It is ideal for long-term monitoring of low-velocity signals. Since the original points are recorded for a long time, the amount of data can be very large; being. As the oscilloscope's DDR memory is limited, real-time sample points are stored in the device's internal flash or on an external U-disk (USB storage device). device) in real time. After completing the acquisition, the user can not only view the sampled points on the oscilloscope, but also analyze the stored data on the computer.

Datalogger_Flow_Chart-3
Datalogger_Record-4

Considerations on the pre-test of the sample logger

The Sample Logger logs the activated channels.

Before you begin, you should:
1. Configure the scale, type and enable the required input channels so anyone can capture the signals you want to record.
2. What is Sample Logger's sample rate based on your settings?
3. How long do you want to record data?
4. Calculate the amount of data needed.
5. Choose to store in internal flash memory or on an external U disk.

 

Logging with the Sample Logger

The steps to log:
1. Enable channels and set channel parameters.
2. Open the Sample Logger and configure the parameters.
3. Starting and stopping the recording In order to reduce the amount of recorded data and speed up the recording, the recording file is recorded in binary format and the file name is *.slg.
The file format is specified in the document "how to extract data from the binary file.docx". Please download this document from an
official SIGLENT website.

Sample_logger_menus-5

Retrieve a logged file in Sample Logger

The Sample Logger can transfer readings from a U disk or flash memory to the oscilloscope for display and analysis. The DDR of the oscilloscope is limited. When the waveform data exceeds the DDR capacity, the U disk is read out during operation; therefore, do not pull out the U disk when analyzing the samples.

1. Get samples. 

 

Sample_logger_menus_recall-6

2. Use the time base and delay controls to adjust the waveform.
3. Use the cursors.

Sample_logger_recall_cursors-7

Analyzing a Sample Logger file on a PC

The recorded file in the flash memory can be exported to a hard disk (USB memory stick) for analysis on the PC.

Sample_logger_export-8

The oscilloscope provides a tool to convert recorded files (.slg) to CSV: File Converter. You can export this tool from the oscilloscope's Save/Recall menu or from the oscilloscope's web page (four-channel X-E models only).

Sample_logger_export_fileconverter-9
Sample_logger_export_fileconverter_webcontrol-10

You can also parse the *.slg file with your own program. The file format is described in the document "how to extract data from the binary file.docx". Please download the document from the official SIGLENT website.

 

Data logger: measurement data logger

The Measure Logger collects readings and displays the history of low-velocity signals. Since the amount of measurement data is relatively small, the data is stored in the oscilloscope's memory for fast processing. After the recording is complete, the data can be saved to internal flash memory or external U disk and imported to PC.

Measurelogger_Flow_Chart-11
Measurelogger_Record-12

Measurement logger considerations before testing

The measurement logger logs the currently selected measurement values.

Before you begin, consider the following:
1. What metrics do you want to log?
2. What sampling interval and duration do you need?

 

Recording with Measure Logger

1. Check the metrics you want to log.

 

measure_logger_step_13

2. Open the Measure Logger and configure the parameters.

 

measure_logger_step2_14

3. Start and stop recording.

measure_logger_step3_15

Analyze, store and retrieve Measure Logger data

1. Use the time base and delay sliders to adjust the waveform.

2. Use the cursors.

Measure_logger_recall_cursors_16

3. Saving and retrieving the recorded data.

Measure_logger_recall_export_17

You can save the recorded data in flash or hard disk. The recorded file is saved in binary format, the file name is *.mlg. The file format is described in the document "how to extract data from the binary file.docx". Please download this document from Siglent official website.

 

Analyzing a Measure Logger file on a PC

For analysis on PC, the logged file in flash can be exported to a U disk. The oscilloscope provides a tool to convert recorded files (.mlg) to CSV: the file converter. You can access this tool from the menu; "Save/Recall" the oscilloscope or export via webpage.

Measure_logger_recall_export_2_18

You can also parse the *.mlg file with your own program. The file format is described in the document "how to extract data from the binary file.docx". Please download the document from the official SIGLENT website. 

Siglent digital oscilloscopes with data logger

SDS1104X-ESDS1104X-E - Back
SDS1104X-E Oscilloscope
SDS1104X-E
The Siglent SDS1104X-E is suitable for a wide range of applications with a bandwidth of 100 MHz & countless software features.

€407.00*
SDS1204X-ESDS1204X-E Back
SDS1204X-E Oscilloscope
SDS1204X-E
The Siglent SDS1204X-E is suitable for a wide range of applications with a bandwidth of 200 MHz & countless software features.

€593.00*
SDS2104X Plus SDS2104X Plus - Back
SDS2104X Plus Oscilloscope
SDS2104X Plus
The SDS2104X Plus oscilloscope from Siglent offers a bandwidth of 100 MHz, as well as 4 analog and 16 digital channels.

€1,199.00*