Description
The TH9201 series are single-phase AC/DC hipot/insulation testers with programmable test sequences for dielectric strength (AC/DC), insulation resistance and open/short check. The TH9201S model supplements this with an in-built 8-channel scanner for parallel test channels, thus increasing the efficiency of multi-point tests.Features
- Four test functions: AC Withstanding Voltage, DC Withstanding Voltage, Insulation Resistance (IR), Open/Short-Check (OS)
- AC output up to 5.00 kV AC / 30 mA, DC output up to 6.00 kV DC / 10 mA with high accuracy ±(1 % reading + 5 digits) and fast discharge function
- Insulation resistance test 0.05 kV-1.00 kV, 0.1 MΩ-10 GΩ (TH9201S up to 10 mA) with automatic range and software control
- Open/short check for contact testing before HV test
- 8-channel matrix scanner integrated in TH9201S for parallel tests (optionally up to 16 channels with external scanners)
- 240×64 dot-matrix LCD for clear display, numeric keypad and rotary knob for intuitive operation
- Programmable test sequences: up to 50 files × 100 steps (500 steps total), memory for start parameters and USB backup
- Fast voltage rise/fall times adjustable (0.1 s-999 s) and waiting times (0.1 s-99.9 s)
- Arc detection, electric wall (leakage current protection), ground current detection, key lock, PASS/FAIL display with acoustic signal
- Comprehensive interfaces: Standard RS232, USB host/device, HANDLER/PLC, SCAN interface; optional GPIB and LAN/Modbus/SCPI
- Active safety functions: Auto-discharge after end of test (DC), protection if exceeded, automatic voltage regulation (±(1 %+10 V))
- Robust design, compact dimensions (340×120×450 mm), temperature range 0-40 °C, humidity ≤90 % RH
Applications
- Automated test benches in production lines (e.g. household appliances, lights, electronic modules)
- Safety tests on transformers, motors, HV components, circuit boards and insulation materials
- Laboratory and development environments for design verification and quality assurance
- Multi-channel tests on components with multiple test points (TH9201S with internal 8-channel scanner)
- Research and training: High-voltage insulation and breakdown tests
- Measuring and testing systems with external control via PLC/RS232/USB/GPIB for automated processes
Technical Data
| Model | ![]() TH9201 | ![]() TH9201S |
|---|---|---|
| Test modes | ACW, DCW, IR, OS | ACW, DCW, IR, OS with internal 8-channel scanner |
| AC voltage range | 0.05-5.00 kV AC (50/60 Hz) | 0.05-5.00 kV AC (50/60 Hz) |
| AC current range | 0.01-30.00 mA; short-circuit >60 mA (V>500 V) | 0.01-30.00 mA; short circuit >60 mA (V>500 V) |
| AC output power | up to 150 VA (5 kV/30 mA) | up to 150 VA (5 kV/30 mA) |
| DC voltage range | 0.05-6.00 kV DC | 0.05-6.00 kV DC |
| DC current range | 0.1 µA-10.00 mA | 0.1 µA-10.00 mA |
| DC output power | up to 50 VA (5 kV/10 mA) | up to 50 VA (5 kV/10 mA) |
| IR voltage range | 0.05-1.00 kV | 0.05-1.00 kV |
| IR resistance range | 0.1 MΩ-10 GΩ (max. 10 mA) | 0.1 MΩ-10 GΩ (Max. 10 mA) |
| IR output power | up to 10 VA (1 kV/10 mA) | up to 10 VA (1 kV/10 mA) |
| Waveform | Sine wave (AC) | |
| Dielectric strength accuracy | ±(1 % reading + 5 digits) | ±(1 % reading + 5 digits) |
| Accuracy IR voltage | ±(1 % reading + 2 V) | ±(1 % reading + 2 V) |
| Voltage resolution | 1 V | 1 V |
| Discharge | Auto-discharge after end of test (DC) | Auto-discharge after end of test (DC) |
| Voltage rise/fall | 0.1-999 s adjustable | 0.1-999 s adjustable |
| Waiting time (DC) | 0,1-99,9 s | 0,1-99,9 s |
| Memory | 50 files × 100 steps (500 steps in total) | 50 files × 100 steps (500 steps total) |
| Operating display | 240×64 dot-matrix LCD | 240×64 dot-matrix LCD |
| Interfaces | RS232, USB host/device, HANDLER/PLC, SCAN; optional GPIB | RS232, USB host/device, HANDLER/PLC, SCAN; optional GPIB |
| Safety functions | Arc detection, electric wall, ground current detection, key lock, OVP/OCP/OPP/OTP | Arc detection, electric wall, ground current detection, key lock, OVP/OCP/OPP/OTP |
| Environment | 0-40 °C, ≤90 % RH | 0-40 °C, ≤90 % RH |
| Mains voltage | 90-110 V / 108-132 V / 198-242 V / 216-260 V AC, 47.5-63 Hz | 90-110 V / 108-132 V / 198-242 V / 216-260 V AC, 47.5-63 Hz |
| Power consumption | ≤500 VA | ≤500 VA |
| Dimensions (W×H×D) | 340×120×450 mm | 340×120×450 mm |
| Weight | ≈15 kg | ≈15 kg |
| Scanner | Internal 8-channel matrix (parallel tests); expandable to 16 channels external | |
| Further specifications | Analogue as TH9201; additional scanner functionality |










.jpg)
.jpg)
.jpg)
.jpg)