Description
The Techmize TH2840A is a precision bench-top LCR meter for component evaluation in R&D, quality control and production. It covers 20 Hz to 500 kHz with 0.05% basic accuracy and measures at up to 1800 readings per second, without relay switching time.
A 20 Vrms test level, a built-in ±40 V / ±100 mA DC bias and a 2 A current source let you test capacitors, inductors and transformers under conditions close to their real operating point. Results, sweeps and limits are set up on a 10.1" touch screen.
Key specifications
- Test frequency: 20 Hz – 500 kHz
- Basic accuracy: 0.05%
- Speed: up to 1800 meas/s (0.56 ms, > 10 kHz)
- Test level: 5 mVrms – 20 Vrms; 50 µArms – 100 mArms
- DC bias: 0 to ±40 V, 0 to ±100 mA; built-in 0 – 2 A current source
- Display: 10.1" capacitive touch screen, 1280 × 800
- Interfaces: LAN, USB Device, 2 × USB Host, RS232C, Handler, External DC bias control
Features
- High throughput: Fast+ mode takes a reading in 0.56 ms (1800 meas/s above 10 kHz). Range changes need no relay switching time.
- Real operating conditions: test levels up to 20 Vrms or 100 mArms, constant level (ALC) at the DUT, and ±40 V / ±100 mA DC bias.
- 2 A current source: built-in DC current source for inductance measurement under DC load.
- Four parameters at once: choose any four of Cp/Cs, Lp/Ls, Rp/Rs, |Z|, |Y|, X, G, B, θ, D, Q, Vac, Iac and Rdc.
- Graphic sweep: plot up to four parameters over frequency, AC level or DC bias, with 51 to 801 points, linear or log axes and cursor read-out.
- 201-point list sweep: per-point averaging, delay, limits and sorting – suited to automated multi-condition tests.
- Production sorting: 10 bins plus PASS/FAIL, up to four limit parameters per bin, bin counter and front-panel PASS/FAIL LEDs. Handler output for automated lines.
- Easy to use:1" touch screen, Linux-based system and built-in help. Mouse and keyboard can be connected via USB.
- Storage: 100 MB internal memory for setup files, plus USB stick storage for setups, screenshots and record files.
- Setup migration: PC software converts setup files from earlier Tonghui models.
Applications
- Passive components: capacitors, inductors, magnetic cores, resistors, piezoelectric devices, transformers, chip and network components
- Semiconductor devices: C-Vdc characteristics of varactors; parasitic parameters of transistors, ICs and LED driver ICs
- Other components: PCBs, relays, switches, cables and batteries
- Materials: dielectric constant and loss of plastics and ceramics; permeability and loss of ferrite and amorphous materials (with a suitable fixture)
- Production: incoming inspection, final test and automated component sorting
Is the TH2840A right for you?
Choose the TH2840A for precision component testing up to 500 kHz. If you need to characterise components or materials up to 2 MHz, see the Techmize TH2840B.
Technical Data
| Specification | ![]() TH2840A | ![]() TH2840B |
|---|---|---|
| Test frequency | 20 Hz – 500 kHz | 20 Hz – 2 MHz |
| Frequency resolution (highest band) | 1 Hz (100 – 500 kHz) | 10 Hz (1 – 2 MHz) |
| AC test voltage | 5 mVrms – 20 Vrms | 5 mVrms – 20 Vrms (≤ 1 MHz) 5 mVrms – 15 Vrms (> 1 MHz) |
| Basic accuracy | 0.05% | 0.05% |
| Frequency accuracy | 0.01% | 0.01% |
| Lowest frequency resolution | 0.1 mHz (20 – 99.9999 Hz) | 0.1 mHz (20 – 99.9999 Hz) |
| Measurement speed | Fast+ 0.56 ms (1800 meas/s, > 10 kHz) Fast 3.3 ms / Medium 90 ms / Slow 220 ms | Fast+ 0.56 ms (1800 meas/s, > 10 kHz) Fast 3.3 ms / Medium 90 ms / Slow 220 ms |
| AC voltage accuracy | ±(10% × set + 2 mV) ≤ 2 Vrms ±(10% × set + 5 mV) > 2 Vrms | ±(10% × set + 2 mV) ≤ 2 Vrms ±(10% × set + 5 mV) > 2 Vrms |
| AC test current | 50 µArms – 100 mArms | 50 µArms – 100 mArms |
| Constant level (ALC) | Voltage and current | Voltage and current |
| Output impedance | 30 Ω ±4% / 100 Ω ±2% @ 1 kHz | 30 Ω ±4% / 100 Ω ±2% @ 1 kHz |
| Rdc test voltage | 100 mV – 20 V | 100 mV – 20 V |
| DC bias voltage | 0 to ±40 V | 0 to ±40 V |
| DC bias current | 0 to ±100 mA | 0 to ±100 mA |
| Built-in current source | 0 – 2 A, 1 mA resolution | 0 – 2 A, 1 mA resolution |
| AC parameters | Cp/Cs, Lp/Ls, Rp/Rs, |Z|, |Y|, R, X, G, B, θ, D, Q, Vac, Iac | Cp/Cs, Lp/Ls, Rp/Rs, |Z|, |Y|, R, X, G, B, θ, D, Q, Vac, Iac |
| DC parameters | Rdc, Vdc, Idc | Rdc, Vdc, Idc |
| Parameters on screen | 4, freely selectable | 4, freely selectable |
| Test terminals | Four-terminal pair | Four-terminal pair |
| Correction | Open, Short, Load | Open, Short, Load |
| List sweep | 201 points | 201 points |
| Graphic sweep | 51 – 801 points, up to 4 traces | 51 – 801 points, up to 4 traces |
| Comparator | 10 bins + PASS/FAIL, bin count 0 – 99999 | 10 bins + PASS/FAIL, bin count 0 – 99999 |
| Display | 10.1" touch screen, 1280 × 800 | 10.1" touch screen, 1280 × 800 |
| Internal storage | Approx. 100 MB setup files | Approx. 100 MB setup files |
| Interfaces | LAN, USB Device, 2 × USB Host, RS232C, Handler, Ext. DC bias control | LAN, USB Device, 2 × USB Host, RS232C, Handler, Ext. DC bias control |
| Power supply | 100 – 120 / 198 – 242 VAC, 47 – 63 Hz | 100 – 120 / 198 – 242 VAC, 47 – 63 Hz |
| Warm-up time | 60 min | 60 min |
| Dimensions (W × H × D) | 430 × 177 × 265 mm | 430 × 177 × 265 mm |
| Weight | 11 kg | 11 kg |
Documentation
eVision Praxis











.jpg)
.jpg)
.jpg)
.jpg)