Description
The Techmize TH2840B is a precision LCR meter for engineers who need to characterise components and materials well beyond audio frequencies. It measures from 20 Hz to 2 MHz with 0.05% basic accuracy, so you can follow capacitors, inductors and magnetic cores into the range of modern switching converters.
With a 20 Vrms test level (15 Vrms above 1 MHz), a ±40 V / ±100 mA DC bias, a built-in 2 A current source and graphic sweep, the TH2840B shows how impedance changes with frequency, signal level and bias, not just a single reading.
Key specifications
- Test frequency: 20 Hz – 2 MHz
- Basic accuracy: 0.05%
- Speed: up to 1800 meas/s (0.56 ms, > 10 kHz)
- Test level: 5 mVrms – 20 Vrms (≤ 1 MHz), up to 15 Vrms above 1 MHz; 50 µArms – 100 mArms
- DC bias: 0 to ±40 V, 0 to ±100 mA; built-in 0 – 2 A current source
- Display: 10.1" capacitive touch screen, 1280 × 800
- Interfaces: LAN, USB Device, 2 × USB Host, RS232C, Handler, External DC bias control
Features
- 2 MHz frequency range: characterise components and magnetics at switching frequencies, with frequency resolution down to 0.1 mHz at the low end.
- Graphic sweep analysis: plot up to four parameters against frequency, AC level or DC bias, with 51 to 801 points, linear or log axes, 1, 2 or 4 screen layouts and cursor read-out.
- C-V and bias characterisation: ±40 V / ±100 mA DC bias and a 2 A current source for varactor C-V curves and inductance under DC load. External DC bias control supports the TH1778A.
- High test level: up to 20 Vrms (15 Vrms above 1 MHz) or 100 mArms, with constant-level ALC at the DUT.
- Fast and repeatable: Fast+ mode at 0.56 ms per reading (1800 meas/s above 10 kHz), with no relay switching time.
- 201-point list sweep: per-point averaging, delay, limits and sorting for multi-condition tests.
- Sorting and automation: 10 bins plus PASS/FAIL, Handler output, LAN, USB (USBTMC) and RS232C for test systems.
- Touch operation:1" 1280 × 800 touch screen, four parameters on screen, Linux-based system with built-in help.
- Storage: 100 MB internal setup storage and USB stick storage for setups, screenshots and record files.
Applications
- Magnetic materials: permeability and loss angle of ferrite and amorphous materials
- Dielectric materials: dielectric constant and loss angle of plastics and ceramics
- Semiconductor materials and devices: dielectric constant, conductivity and C-V characteristics; varactor C-Vdc; parasitic parameters of transistors, ICs and LED driver ICs
- Liquid crystal cells: dielectric constant, elastic constant and C-V characteristics
- Passive components: capacitors, inductors, cores, resistors, piezoelectric devices, transformers, chip and network components
- Other components: PCBs, relays, switches, cables and batteries
Is the TH2840B right for you?
Choose the TH2840B when your work extends above 500 kHz or into material characterisation. For component testing up to 500 kHz, the Techmize TH2840A offers the same platform.
Technical Data
| Specification | ![]() TH2840A | ![]() TH2840B |
|---|---|---|
| Test frequency | 20 Hz – 500 kHz | 20 Hz – 2 MHz |
| Frequency resolution (highest band) | 1 Hz (100 – 500 kHz) | 10 Hz (1 – 2 MHz) |
| AC test voltage | 5 mVrms – 20 Vrms | 5 mVrms – 20 Vrms (≤ 1 MHz) 5 mVrms – 15 Vrms (> 1 MHz) |
| Basic accuracy | 0.05% | 0.05% |
| Frequency accuracy | 0.01% | 0.01% |
| Lowest frequency resolution | 0.1 mHz (20 – 99.9999 Hz) | 0.1 mHz (20 – 99.9999 Hz) |
| Measurement speed | Fast+ 0.56 ms (1800 meas/s, > 10 kHz) Fast 3.3 ms / Medium 90 ms / Slow 220 ms | Fast+ 0.56 ms (1800 meas/s, > 10 kHz) Fast 3.3 ms / Medium 90 ms / Slow 220 ms |
| AC voltage accuracy | ±(10% × set + 2 mV) ≤ 2 Vrms ±(10% × set + 5 mV) > 2 Vrms | ±(10% × set + 2 mV) ≤ 2 Vrms ±(10% × set + 5 mV) > 2 Vrms |
| AC test current | 50 µArms – 100 mArms | 50 µArms – 100 mArms |
| Constant level (ALC) | Voltage and current | Voltage and current |
| Output impedance | 30 Ω ±4% / 100 Ω ±2% @ 1 kHz | 30 Ω ±4% / 100 Ω ±2% @ 1 kHz |
| Rdc test voltage | 100 mV – 20 V | 100 mV – 20 V |
| DC bias voltage | 0 to ±40 V | 0 to ±40 V |
| DC bias current | 0 to ±100 mA | 0 to ±100 mA |
| Built-in current source | 0 – 2 A, 1 mA resolution | 0 – 2 A, 1 mA resolution |
| AC parameters | Cp/Cs, Lp/Ls, Rp/Rs, |Z|, |Y|, R, X, G, B, θ, D, Q, Vac, Iac | Cp/Cs, Lp/Ls, Rp/Rs, |Z|, |Y|, R, X, G, B, θ, D, Q, Vac, Iac |
| DC parameters | Rdc, Vdc, Idc | Rdc, Vdc, Idc |
| Parameters on screen | 4, freely selectable | 4, freely selectable |
| Test terminals | Four-terminal pair | Four-terminal pair |
| Correction | Open, Short, Load | Open, Short, Load |
| List sweep | 201 points | 201 points |
| Graphic sweep | 51 – 801 points, up to 4 traces | 51 – 801 points, up to 4 traces |
| Comparator | 10 bins + PASS/FAIL, bin count 0 – 99999 | 10 bins + PASS/FAIL, bin count 0 – 99999 |
| Display | 10.1" touch screen, 1280 × 800 | 10.1" touch screen, 1280 × 800 |
| Internal storage | Approx. 100 MB setup files | Approx. 100 MB setup files |
| Interfaces | LAN, USB Device, 2 × USB Host, RS232C, Handler, Ext. DC bias control | LAN, USB Device, 2 × USB Host, RS232C, Handler, Ext. DC bias control |
| Power supply | 100 – 120 / 198 – 242 VAC, 47 – 63 Hz | 100 – 120 / 198 – 242 VAC, 47 – 63 Hz |
| Warm-up time | 60 min | 60 min |
| Dimensions (W × H × D) | 430 × 177 × 265 mm | 430 × 177 × 265 mm |
| Weight | 11 kg | 11 kg |